User settings:
The commands
LDW
and
LWT
work like the commands
SZA
and
SFA
. It is possible with these commands to
make a system-specific calibration (e.g. of a scale), without changing the transducer settings performed with
SZA
and
SFA
.
The user-specific settings LDW/LWT result as an ASCII output within the range 0... 1000000. This characteristic
is converted into the
NOV
by means of the parameters
Example:
NOV 0;
User setting 0...1000000 (ASCII signal
with 0 = unloaded scale, 1000000 at nominal load
NOV 4000; User setting 0...4000
With output 0 at load =0 or 4000 at nominal load
Attention:
In an adjustment with SZA and SFA, the parameters for LDW and LWT are reset (default: LDW=0 and
LWT=1000000) .There is the option to determine the characteristic with SZA and SFA at any time by entering
LDW0 and LWT1000000. Before setting your new characteristic, enter LDW0,LWT1e6, and NOV0.