AD104-R2, AD104-R5
33 / 75
ba_aed104_8_e.doc
LWT
Load Cell Weight
(nominal value of the user characteristic)
Range:
0...1,599999e6
Factory setting:
1000000
Response time:
<15ms...4.2s
Parameters:
1 (0)
Password protection:
yes
Parameter protect.:
with input
Take the full scale value of user characteristic with LWT command
Input:
LWT;
(Response time <4,2s):
Using this command, the transducer electronic system measures an input signal
between ±2.5mV/V ( i.e.= measured value at scale nominal load), stores this
measured value as the nominal value and calculates the new user characteristic
together with the previously entered value for
LDW
. The values for
SZA
and
SFA
are not changed.
Individual input of the full scale value of the user characteristic via LWT:
Input:
LWT<Nominal value>;
(Reaktionszeit<1,5s):
Instead of causing the AED to measure the applied signal, the value for the end value of the
user characteristic is entered here directly and, in combination with the previously entered
value for
LWT
, used in calculating the user characteristic.
Query:
LWT?;
(Response time <15ms)
Effect:
Output the the value for the user characteristic in ±7 digits (e.g. -0000345CRLF) of the
nominal value or the loaded transducer (nominal load). The value is not converted via
NOV
.
Important Note
An input or measurement of the factory characteristic with
SZA/SFA
resets the user characteristic to the default
values
LDW
=0,
LWT
=1000000.
Procedure for entering the user characteristic:
1.
Enter password by means of command SPW.
2.
NOV 0;
output (scaling off).
3.
Adjust the ASF filter such that a maximally quiescent display is effected
4.
Scale at no load , wait for standstill