34 / 75
ba_aed104_8_e.doc
5.
NOV
Nominal Value
(Resolution of the user characteristic)
Range:
0...1,599999e6
Factory setting:
0 (=deactivated)
Response time:
<15ms
Parameters:
1
Password protection:
yes
Parameter protect.:
with command
TDD1
Input:
NOV<Wert>;
Query:
NOV?;
(Response time <15ms)
Effect:
The value stored in the AED is output with 7 digits complete with sign ( e.g.
0001000CRLF 0000345crlf ).
The
NOV
value is used for scaling the measured value. At
NOV
=0 this output scaling is deactivated. The ASCII
measured value output is scaled to 1000000 at the factory. If a measured value output of 2000 digit at nominal
load is required, then use this command to set the nominal value
NOV2000
. The input parameters or tara
values are not changed by this scaling.
Measured value output format at nominal
load
NOV=0
NOV>0
2 Byte binary
20000
NOV
value
4 Byte binary
5120000
NOV
value
ASCII
1000000
NOV
value
With the 2 byte binary type of output, the
NOV
value must be < 30000. Otherwise the measured value will be
output complete with overflow or underflow ( 7fff
H
or 8000
H
; H: Hexadecimal). With
NOV30000
, the overload
range is only approx. 2700 digits.