1 Intended
use
The digital sensor electronic units AD104 belong to the family of AED components which digitally condition and
network as bus-capable signals of mechanical measured value transducers. The objective of these components
is the digitization and conditioning of the measuring signals directly at the transducer. The AD104 and the
transducer (load cell) form a unit and cannot be replaced separately (transducer calibration of the measurement
chain with SZA/SFA is necessary).
As transducers, calibrated load cells or force transducers ( adjusted in TCZ,TCS, and zero point) can be used.
The measuring amplifier boards AD104 have different interface connections which are produced in the factory
by corresponding assembly:
AD104-Type
Interface
Interface
Connect.
Bus mode
Cable length
external
Trigger
AD104-R2
Asynchronous,
serial
RS-232 duplex
no
< 15m
yes
AD104-R5
Asynchronous,
serial
RS-485 -4-wire,
full duplex
yes
< 500m
yes
The transducer electronic units AD104 are also abbreviated with AED in the following text.
Old type: AD104-R4
Æ
new type: is AD104-R5 with external trigger
New type: AD104-R2 with external trigger and RS232 interface.
2 Characteristic
features
•
Operating voltage 5.6V...15V DC
•
4 wire Interface for a full bridge sensor, nominal input range
±
2 mV/V, maximal input range
±
2.6 mV/V
•
Serial interface RS-485 (bus mode) or through RS-232 interface (point - to - point mode)
•
Digital filtering, choice of the output speed and scaling of the measured signal
•
Separated calibration of transducer and application characteristic
•
Storage of the parameters nonvolatile
•
All settings are made through the serial interface
•
Automatic zero tracking (1d/s,
±
2%)
•
Automatic intial zero setting (
±
2%...
±
20%)
•
Trigger functions (level or external trigger)