LDW
Load Cell Dead Weight
(Zero point of the user characteristic)
Range:
0...1.599999e6
Factory setting:
0
Response time:
<15ms...4.2s
Parameters:
1
Password protection:
yes
Parameter protect.:
after input of
LWT
Take the zero point of user characteristic with LDW
Input:
LDW;
(Response time: <4.2s)
1.
The scale is unloaded.
Take the zero point with
LDW;
command.
Using this command, the transducer electronic system measures an input signal between ±2.5mV/V ( i.e.
=measured value for scale in no load condition), stores the measured value, but only starts the calculation
of the new user characteristic after entering the parameter for
LWT
.
Individual input of the zero point of the user characteristic via LDW
Input:
LDW<Zero point>
(Response time: <15ms)
1.
Insert the value for the zero point of the scale via
LDW<Zero point>
command. The entered value is
stored, but accounted for only after entry of the parameter for
LWT.
Query:
LDW?;
(Response time<15ms)
Effect:
Output in ±7 digits (e.g. -0915345crlf) of the zero point value. The value is not converted via
NOV
.