AD104-R2, AD104-R5
55 / 75
ba_aed104_8_e.doc
ZSE
Zero Setting
(Automatic zero setting)
Range:
0...4
Factory setting:
0
Response time:
<0.1s
Parameters:
1
Password protection:
none
Parameter protect.:
with command
TDD1
Input:
ZSE(0...4);
0 - Zero setting deactivated,
1 – Zero setting range + 2% of NOV value
2 - Zero setting range + 5% of NOV value
3 - Zero setting range + 10% of NOV value
4 - Zero setting range + 20% of NOV value
Query:
ZSE?;
Response:
0...4
Function:
After voltage switch-on, or following a RESET, or after the RES command, and
on expiry of a 2.5s delay period, intial zero setting will be executed within the
selected range at a standstill. If there is no standstill, or if the gross value is
outside the selected limits, there will be no zero setting. The internal zero
memory is always deleted before any automatic zero setting. If the gross value
at standstill is within the selected range, this gross value will be stored in the
zero memory. The zero memory cannot be read out. Scale standstill is fixed at
1d/second. The unit ‘d’ (digit) refers to the nominal value (NOV). If the nominal
value is deactivated (NOV=0) or NOV value > 10000d, then there will be a
standstill monitoring relative to a nominal value of 10000d.
If the NOV value is set to <100d, then there will be a standstill monitoring in
accordance with a nominal value of 100d.