8-22
TABLE 8-6
AUTOMATIC COLLIMATOR SPECIFICATIONS
Parameter
Specification
Collimator
Digital version with automatic formats collimation system and light
localizer for rectangular collimation with linear light localizer and rails
for secondary filters
Operation Mode
Automatic
Manual
Dimensions
349 L x 237W x 181 H (mm)
Weight
9.8 +/– 0.5 (kg)
Aperture Angle
28° / 28°
Angle of Rotation around
central beam axis
+/– 45°
Maximum field Size with
SID = 90cm
43cm x 43cm
Maximum field Size with
SID = 100cm
3cm x 3cm
Power Supply:
DC Input Voltage (V)
Max. DC Input Current (A)
ACV Input Voltage (V)
Mac AC Input Current
24 +/– 10%
1.5
28 ... 40 +/– 10%
10
Halogen Lamp
24V~/ 150W / Part Number: 8375545 G2107
Only OEM Siemens parts may be used as replacement parts!
Operation Temperatures
10°C < to > 40°C
Transport and Storage
Temperatures
– 40°C to +70°C
Inherent Filtration
EN 60601–1–3 Standard
1.0 mm Al at 70 kV
Leakage Filtration
EN 60601–1–3 Standard
< 0.5 mGy/h