8-4
The vertical load is balanced by a spring counterpoise system within the carriage.
The counterpoise system is equipped with a
safety
locking feature to prevent the
tube unit from falling in the event of spring or main cable failure. Adding an acces-
sory such as a collimator extension cylinder may cause the suspension to be
slightly out of balance.
ILLUSTRATION 8-3
TELESCOPIC COLUMN AND CARRIAGE
1-5Use of Longitudinal and Transverse
Detents
The suspension is programmed with Longitudinal and Lateral detents to automati-
cally apply the locks and signal the collimator when the tube is positioned at spe-
cific SID’s, for vertical table or wall stand procedures. For table, lateral detent is
set at the table lateral center line. For wall stand, lateral detent is set at the wall
stand panel lateral center line, and the longitudinal detent is set at SID 180cm(72
inches) and 100cm(40 inches). Selection is made at the time of installation and is
usually with the focal spot over the longitudinal table and wall stand center line.
The locks are actuated through electrical detents and the switch marked
DETENT. Depress the “DETENT” switch and the “LAT” switch to index the focal
spot over the longitudinal table center line. Depress and momentarily hold the
“LONG” switch (with the “DETENT” switch still depressed) to index the focal spot
at one of the pre–selected SID’s. When reaching the approximate position, motion
should be slow to avoid passing over the detent bracket. To move out of the
indexed positions, depress the “DETENT” switch a second time.
Collimator accessory weight may not exceed 4.5 kg (10 pounds). Use
special care when using such an accessory since the tube unit will
tend to descend when the vertical lock is released.
TUBE SUPPORT
USER
INTERFACE
TELESCOPING
COLUMN
CARRIAGE
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).