7-34
3-2Precision 500D Tables, Tubes and Collimators
3-2-1
X–Ray Tube Specification
The following X-Ray Tubes may be used with Intelligent Digital Device and Preci-
sion 500D Table (Compatible with fluoro collimator). Refer to Appendix A, Maxiray
100 Tube Units, for further information about these X–Ray Tubes.
TABLE 7-10
X–RAY TUBE COMPATIBILITIES FOR IDD/PRECISION 500D TABLE
TABLE 7-11
LEAKAGE TECHNIQUE FACTORS
TABLE 7-12
MINIMUM INHERENT FILTRATION
NOTE:
Radiopacity of the top and sub-panel is less than 1 mm aluminum equivalent at
100 kVp
X–RAY TUBE
TARGET ANGLE
FOCAL SPOT
POWER
SHIM
TYPICAL APPLICATIONS
MX 100 FL
12.5°
0.6 mm/1.25 mm
32/100 KW
2.1°
DF TEMPORAL
TUBE UNIT
LEAKAGE TECHNIC
FACTORS
kVp
mA
MX 100–FL 0.6–1.25 12.5°
46–155500G18 (Housing) 46–155318G36 (Insert)
150
5
COMPONENT
MINIMUM INHERENT FILTRATION
TUBE UNIT
MX 100–FL 0.6–1.25 12.5°
46–155500G18 (Housing) 46–155318G33 (Insert)
Equivalent to 1.1 mm aluminum at 150 kVp
COLLIMATORS
• RAD Overhead
AL01C
1.9 mm AI / 70kV
• Fluoro Undertable
AF01A
1.5 mm AI / 70kV