6-34
The Length line (actual line or control points) will be included on the film output of
the Stenosis data. If any part of the Length line extends beyond the black circle
mask surrounding the image, it will not be displayed on the film.
18-8Show Data
Select this feature button to display the results for any stenosis measurements
functions recently performed. This window can be moved on the image screen by
selecting the title bar and dragging it to a different area on the screen.
ILLUSTRATION 6-22
RESULTS SCREEN
The
Vessel Diameter
graph represents the distance between the vessel edges. This is only available if the Manual
Boundary process was performed.
Two of the three vertical lines represent the edges of the vessel, and the third indicates the minimum width within
the traced area of the vessel. All three lines are adjustable. You will notice that as the minimum distance line is
moved, the stenosis indicator shown on the image is moved accordingly, and the data within this screen is recalcu-
lated.
Vessel Results:
These measurements are based on the vessel edge traces
entered during the Manual Boundary procedure.
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie
CDL).