5-15
2-2Fluoro Acquisition Screen
ILLUSTRATION 5-15
FLUORO ACQUISITION SCREEN
Selections
•
This screen will display record factors as well as fluoro factors.
•
The top portion of the screen is dedicated for record factors.
•
You may change your contrast type for the trajectories to compensate techni-
cal parameters. However it is recommended to use the appropriate contrast
per procedure.
•
The other option is frames per second. You may select single, 1, 2, 3, 4, 5, 6
or 7.5 maximum.
•
The lower portion of the screen is dedicated to fluoro factors.
•
You may select the image intensifier size. (This is system dependent.)
•
The Field of View options are 16, 12, 9, 6 or 12, 9, 6, 4.5 inch (40, 32, 22,
16 cm or 32, 22, 12, 9 cm) modes.
Pediatric Dose Selection (optional)
- Provides specialized image acquisition
parameters for small pediatric patients. This special purchase option provides
dedicated and tailored dose trajectories for infants and small children.
•
The dose can be limited to one of three rates: the standard 10 R/minute grid
in or 5R/minute input dose (grid in or out).
•
Provides the optimum mix between minimal dose and maximum image qual-
ity.
•
The pediatric mode is not available in Germany or New Zealand due to exist-
ing low dose regulations.
Pulsed Fluoro (option) -
The system works with 15, 7.5 and 3.75 frames per sec-
ond. After selecting, the rates can be changed at the IDD.
FOR
TRAINING
PURPOSES
ONLY!
NOTE:
Once
downloaded,
this
document
is
UNCONTROLLED,
and
therefore
may
not
be
the
latest
revision.
Always
confirm
revision
status
against
a
validated
source
(ie