11-4
11-4
Generator Ouput Power:
TABLE 11-5
JEDI GENERATOR POWER
The JEDI can deliver the power specified in this output for at least 100ms,
depending on the X–ray tube used. Refer to the tube rating charts supplied with
the x–ray tube.
In accordance with IEC requirements,
Time Product.
TABLE 11-6
NOMINAL ELECTRIC POWER (1000KV–0.1S):
TABLE 11-7
PARAMETER SELECTION RANGES OF CONSOLE READOUTS
kV
65kW
80kW
40–80
800 mA
1000 mA
100
630 mA
800 mA
125
500 mA
630 mA
150
400 mA
500 mA
kW
k
65
80
I (max)
mA
630
800
All Modes
Selectable kV
40–150
Decimal scale
steps of 1kV
Jump of 10kV
Renard scale
40, 41, 42, 44, 46, 48, 50,
52, 55, 57, 60, 63, 66, 70,
73, 77, 81, 85, 90, 96, 102,
109, 117, 125, 133, 141,
150
The next kV displayed is
the third Renard step value
following the initial display
value of kV.