5-52
3-2-6
Retrieve
When Retrieve is selected a pop–up window appears. This function allows the
user to retrieve a protocol database that has been saved onto a CD.
ILLUSTRATION 5-52
PROTOCOL RETRIEVE
When the Protocol Retrieve begins, the system looks for a protocol database on
the CD that corresponds with the system configuration.
NOTE:
Initiating a Retrieve causes changes in the protocol database only. No other
areas of the system (i.e. system configuration) is affected.
NOTE:
Each retrieved Protocol is checked for compatibility with the system configu-
ration.
When a retrieved Protocol is found to be incompatible, the technique parameters
are modified by the system to create a valid technique. In the case of a tube not
being available the technique is removed. If it is a smaller x–ray tube the system
will drop the mA value. If the system is not configured for pulse fluoro and a proto-
col being retrieved has pulse fluoro the fluoro rate is changed to continuous.
At the end of the retrieve the user is notified of the changes made to the new data-
base. The user then is able to accept or reject any of the changes. When the user
selects “Accept” the new database replaces the existing one. If the user chooses
“Cancel” the old database remains. Refer to the following screen illustration.