7-33
3-1-1
Precision 500D Table Electrical
Requirements
Electrical Data: Precision 500D Table with Intelligent Digital Device
Ratings: 240VAC, 20 amps, Single phase, 50/60 Hz
Power Supply Requirements:
Generator power unit provides 240VAC, single phase, 50/60 Hz*,
20 amp branch circuit plus ground
Operating Temperature
50 to 100F (+10° to +38°C)
Relative Humidity
20% to 70% (non–condensing)
Altitude
8000 ft. (2438 m) above sea level max.
Storage or Shipping
Temperature
–40° to 140°F (–40°C to 60°C)
Relative Humidity
20% to 80% (non–condensing)