11-6
TABLE 11-9
ACCURACY OF CONSOLE READOUTS
NOTE:
In RAD mode, kVp accuracy is ensured after 10ms.
Display before exposure
Conditions parameter
mode Basic
Accuracy
Radiography
Exposure
Time
<= 40
kVp
all
+/– (3%, +2 kVp)
mA
all
+/– (10%, +1 mA)
ms
3 point timer
+/–(5%, +0.5ms)
ms
AEC (2/3 point)
+/– 5%
mAs
AEC (2/3 point)
+/– 10%
mA
AEC (3 point)
+/– 15%
Radiography
Exposure
Time
> 40 ms
kVp
all
+/– (3%, +2 kVp)
mA
all
+/– (5%, +1 mA)
ms
3 point timer
+/–(5%, +0.5ms)
ms
AEC (2/3 point)
+/– 5%
mAs
AEC (2/3 point)
+/–(5%, +0.1 mAs)
mA
AEC (3 point)
+/– 10%