Section 8-9 - Power Supply Battery Pack Replacement Procedure
Section 8-9
Power Supply Battery Pack Replacement Procedure
8-9-1
Manpower
One person, .5 hour + travel
8-9-2
Tools
Phillips screwdriver
8-9-3
Preparations
(click here to view the cover removal video)
We anticipate,with the exception of a failure, batteries will only need to be replaced once every three
years.
1.) Power Down/Shutdown the system as described in
2.) Remove the left- and/or right-side covers as shown in
3.) Remove the upper rear cover as shown in
.
4.) Remove the lower rear cover as shown in
5.) Remove the Back End Processor as shown in
.
8-9-4
Power Supply Battery Pack Removal Procedure
1.) Remove the cover of the Back End Processor.
2.) Remove the four screws from the power supply battery pack cover plate.
3.) Disconnect the battery pack connection wire.
4.) Remove the power supply battery pack from the Back End Processor.
Figure 8-54 Removing / Installing the Power Supply Battery Pack