GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
8-20
Section 8-4 - Back End Processor Power Supply Replacement
3.) Clip the ty-wraps holding the power supply cables to the side of the power supply.
4.) Mark both ends of the power connectors, as shown in
, on the power
cables coming from the power supply to the:
-
CD
-
MO
-
PC2IO
-
HDD
-
PCI Bird Card (If system has 3D option)
NOTE:
All five power connectors are exactly alike. They should be marked as a reference for cable
routing purposes when the new power supply is installed.
Figure 8-11 LOGIQ 9 Power Supply & Connector Locations
Cut Ty-wraps
Power Supply
Location
PC2IO
CD
MO
CD
HDD