GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
7-62
Section 7-12 - Troubleshooting Trees
Figure 7-53
Error on the monitor display!
Set the Monitor setting to the default.
NO
YES
Do the colors seem
abnormal?
Does the Contrast or the
Brightness adjustment
seem improper?
Adjust Contrast and Brightness.
YES
NO
Does distortion occur on
the monitor? Does the
display area seems
abnormal?
NO
Does the monitor display
nothing?
Replace the monitor.
YES
Check the PCVIC, and AGP video
tape.
YES