GE M
EDICAL
S
YSTEMS
PROPRIETARY
TO
GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
Chapter 3 Installation
3-41
3-15-6
If the System is NOT on a Network
If the system is connected to a network, skip this section.
If the system is
NOT
on a network, a static "dummy" IP-Address must be install in Windows for DICOM
and Daignostics to function.
To install a Static IP-Address in windows:
1.) Boot up the system with the service dongle attached to the USB port.
2.) From the Start Applications dialog box, like the one shown in
, select
MAINTENANCE Mode to enter the Windows desktop.
3.) The Maintenance Password screen will appear. Enter the current Service Password.
4.) Select Exit to Windows.
5.) From the task bar at the bottom of the Touch Panel select START>SETTINGS>NETWORK
DIALUP CONNECTIONS.
, double click on LOCAL AREA CONNECTIONS.
Figure 3-33 Select Maintenace Mode
Figure 3-34 Network and Dialup Connections Window