Section 8-8 - BEP Patient I/O Replacement Procedure (BEP 2.x and above only)
5.) Remove the two screws in the front of the Patient I/O as shown in
6.) Slide the Patient I/O out the front of the BEP.
7.) Remove the interface cable from the back of the Patient I/O.
8-8-5
Patient I/O Replacement Procedure
1.) Attach the interface cable to the back of the new Patient I/O.
2.) Insert the new Patient I/O and cable into the drive bay and secure it in the front with the two screws
previously removed. refer to
.
3.) Attach the one screw to the side of the Patient I/O through the rack. refer to
4.) Attach the Patient I/O interface cable to the BEP IO VIC Tray. Ensure all cables are properly
arranged and secure.