Chapter 7 Diagnostics/Troubleshooting
7-63
•
Monitor does not work
-
Check AC circuit breaker
-
Verify power is present at monitor outlet
-
Check if video is present at backplane
•
Colored streaks appear in image
-
Check for presence of magnetic sources near the monitor
-
Eliminate the sources and then degauss the monitor
-
The monitor should automatically degauss itself each time power is applied if you wait at least
10 seconds before you turn power back ON
•
Prints do not match monitor
-
Verify factory default settings. Chapter 18 in the User Manual calls out suggested settings for
various exam and lighting conditions.
•
Video test patterns are not clear, bright, parallel or square
-
Replace the monitor