GE M
EDICAL
S
YSTEMS
PROPRIETARY TO GE
D
IRECTION
2294854-100, R
EVISION
3
LOGIQ™ 9 P
ROPRIETARY
M
ANUAL
7-42
Section 7-10 - I/O Devices
Test slide pots (TGC) by moving them through their range. Feedback will be returned to the Log area
as shown in
.
The important thing to watch here is the slide pot number that is moving and ensure that it’s value
changes. In
slide pot s0 is changing.
NOTE:
It is important to remember that the trackball and alpha-numeric keyboard are not tested by this
program. Moving the cursor in Windows with trackball would indicate its functionality. PC
Diagnostics in the Common Service Interface will check the alpha-numeric keyboard.
Figure 7-40 Slide Pot Test Results