GE Medical Systems
EAGLE 1000
Page 22
Information Technologies
Servicing Instructions 227 468 01 Rev.G
overrange limit (e.g. for polarization voltages) for differential signals >0.6 V, for signals
referred to RL > 1.0 V
common-mode rejection referred to RL for f<= 50 Hz approx. 80 dB RL referred to
chassis >110 dB
patient leakage current (rms values): under normal conditions <0.005 mA, in fault condi
tion (e.g. patient in contact with mains voltage) <0.020 mA
quiescent input current <40 nA
non-destructive load for electrode connections referred to RL ± 50 V, referred to chas-
sis ± 1000 V
pulse-voltage resistance of all electrode connections and of the RL-connection referred
to chassis (either polarity, e.g. defibrillation, electrocautery) 5000 V
Signal processing
The signal in channel 1 is processed. All specifications refer to a signal free of inter
ference.
Amplification, QRS detection, digitization, display
HR measuring range 15 to 300 bpm, error ± 1 BPM (EC 13)
ECG cut-off frequency >=100 Hz
differential signal range ± 4.5 mV
ECG resolution 0.005 mV
sampling rate 1000/1200 Hz
noise below 0.04 mV
time constant for ECG transfer 0.3 s (0.5 Hz)
HR averaging for displayed value: weighted average from the most recent RR intervals
display update interval: 2 seconds
trigger sensitivity 0.25 mV to 0.30 mV for pulse width of 40 ms
max. rate stabilization of 6 seconds for RR interval change of 95%
alarm delay 4 seconds
asystole alarm 4 seconds
3 signal sizes for monitor display in freeze mode or else on the recorder trace: 5 10
20 mm/mV
1-mV calibration signal on waveform and recorder trace
detection of pacing pulses (these specifications refer to an ECG signal free of inter-
ference)
pulse duration d
p
>approx. 0.2 ms, <40 ms
pace marker independent of polarity
pulse amplitude a
p
±10 to ±700 mV
over/undershoot a
o
±1 mV
time constant t
o
25 to 100 ms
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......