Servicing Instructions 227 468 01 Rev.G
Measurement of non-invasive blood pressure disabled or unreliable
- Cuff unsuitable or defective:
check that the cuff being used has the recommended size,
(children, adult, neonate) and is in good condition
- Tubing between instrument input and PCB is pinched as soon as the instrument casing
is put on, or leaks:
lay tubing appropriately, if necessary, replace
- Component on PCB defective (pump, valve, electronics):
replace PCB
- PCB NBP wire set not plugged correctly or defective:
plug in correctly or replace it
- NBP interface for non-invasive blood pressure on PCB ECG Monitor is defective:
re-
place this PCB
- Signal transmission on PCB ECG Monitor defective:
replace this PCB
No measurement by TEMP
- Temperature probe defective:
check the cable and the plug, if necessary, replace probe
- One of the two control temperatures is incorrect:
from Configuration access the Ser-
vice Mode with F3 and F4 and read the control temperatures displayed. If the upper
value deviates from 38.8 °C by more than 0.1 °C, the lower value from 4 °C by more
than 0.2 °C, it is no longer possible to make an accurate measurement and probe
value output is disabled (> 0.2 °C internal deviation for both values). The PCB
PRESS/TEMP, or PCB TEMP must be replaced
- The socket is not correctly inserted or is defective:
plug in flat strip cable correctly, if
necessary, replace it.
- Power supply from PCB ECG Monitor is too weak:
provided there is no overloading
due to the temperature or Press module, PCB Oximeter, or the ECG circuit, the PCB
ECG Monitor should be replaced
- Signal transmission on PCB ECG Monitor defective:
replace this PCB
No measurement by PRESS
- Pressure transducer or cable defective:
inspect transducer membrane, subject to pres-
sure, check cable, if necessary, replace transducer (or cable, if connectable)
- When using the Lionheart simulator it is not possible to display a pressure signal on
the display, although the socket and power supply are O.K. and plugged in correctly:
replace PCB PRESS/TEMP
- Signal transmission on PCB ECG Monitor defective:
replace this PCB
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......