GE Medical Systems
EAGLE 1000
Page 86
Information Technologies
Servicing Instructions 227 468 01 Rev.G
4. 0 °C
Name
4.1 Service mode tests
Using F3 // F4 call up the service mode from the configuration mode, the latter with F1 //
F5.
Key test with short press, actuate key function by pressing key for about 2 seconds. Quit
service mode with F3 // F4.
Instrument configuration is identified as display is generated on the service screen, do not
quit prematurely.
Key F1 / Print
Printhead test
The entire screen should show
diagonal lines
Key F2 /Dump
Screen printout
In the case of TEMP the control values
are shown.
Key F3 / RS 232 Enables interface to be tested
Key F4 / -
Key F5 / Mano Measurement of a steady pressure in NBP
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......