Information Technologies
Servicing Instructions 227 468 01 Rev.G
Testing connector for measuring patient leakage current
The following wiring of testing connectors is recommended for the technical inspection of
module input type BF (body floating) and type CF (cardiac floating).
The catalog numbers of the Marquette Hellige components are given below.
When making the connectors observe the following configurations:
The diagrams of the plug inserts in this document are always as viewed from the inside of
the connector shell.
- Testing connector for the temperature input
1 x
2-pin jack plug
91530000
1 x
Telephone socket or
91534800
matching isolated plug connection for testing case
The two jack plug poles are short-circuited with one another and connected to the
telephone socket.
- Testing connector for ECG input
1*
Connector
91541378
10*
Contact pins
91541394
1*
Telephone socket or
91534800
matching isolated plug connection for testing case
Short-circuit all 10 contact pins with one another and connect to
telephone socket.
- Testing connector for SaO2 input
1*
Connector
91541871
7*
Contact pins
91541394
1 *
Telephone socket or
91534800
matching isolated plug connection for testing case
6*
10 kohm resistor
92184443
Connect pin 10 directly to the telephone socket.
Connect pins 1,4,5,6,7,9 via 10 kohm resistor to telephone socket.
- Testing connector for PRESS input
1*
Testing connector parts
91541376
7*
1 kohm resistor
92184430
9*
Contact pins
91541394
1*
Telephone socket or
91534800
matching isolated plug connection for testing case
Connect pins 7 and 10 directly to telephone socket.
Connect pins 1,2,3,4,6,8,9 separately via 1 kohm resistor to telepho-
ne socket.
1
2
3
4
5
7
8
9
10
6
10
8
6
3
1
5
2
7
4
9
10
1
4
3
2
6
8
5
9
7
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......