GE Medical Systems
EAGLE 1000
Page 96
Information Technologies
Servicing Instructions 227 468 01 Rev.G
3-4 As described above under offset.
5a) Eagle 1000 without recorder:
In the ECG menu set softkey mm/mV to 10. Measure the 1-mV marker on the
screen. Adjust this amplitude of the square wave edge on the screen using
R517
for
channel 1 and
R530
for channel 2. Provided the adjustment has been carried out
correctly, after actuating Freeze the amplitude of the square wave edges is 10 mm.
5b) Eagle 1000 with recorder:
In the ECG menu set softkey mm/mV to 10. Select Print. Adjust 10 mm amplitu-
de of the square wave edge using R517 for channel 1 and R530 for channel 2 in the
recording.
DISPLAY ADJUST
Adjustment procedure for HOSIDEN display 930 117 17 or 2009071-001
CONTRAST ADJUSTMENT
Select DISPLAY in the menu (can be called up with the Next Menu key). Press Con-
trast- until the darkest setting has been attained. Using R776 adjust the contrast so that
the display image is still just visible.
DISPLAY CONTRAST SETTING
Keep C depressed until the background noise viewed on the screen vertically
disappears. When the contrast setting is increased further this leads to bloom on the dis-
play.
Adjustment procedure for HITACHI SP14Q002-A1 display 2003727-001 or -004
Press F1 and F5 together and write down the actual monitor defaults. Set status for
DEFAULTS to on, then press ENTER. Now press EXIT and APPLY and SAVE.
Switch unit off and on again, now you have the manufacturer default contrast of the
display. With R776, adjust for best contrast when looking straight to the front to the display.
After this is done, check the monitor defaults with the written down defaults and change
them back if necessary.
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......