GE Medical Systems
EAGLE 1000
Page 57
Information Technologies
Servicing Instructions
227 468 01 Rev. G
PIN CHART
J1
Pin 1
AC Live
Pin 2
AC Neutral
J 2
Pin 1
+Vout
Pin 2
+Vout
Pin 3
+Vout
Pin 4
Return
Pin 5
Return
Pin 6
Return
P1
Pin 1
Safety Earth Ground
2.2.13 PCB Power Supply 930 117 41
2.2.13.1 Function and pin assignment
The PCB Power Supply comprises a primary clock-rated wide-range power supply unit
with an output voltage separated by a transformer. Thus a voltage selector is not
necessary. The power supply unit fullfills leakage current requirements according to IEC
601 and UL 544. It is tested according to IEC, UL and CSA.
AC = 95 to 240 V AC rated voltage (85 to 264 operating voltage)
Frequency 49 to 65 Hz
Dielectric strength 1500 V AC
+Vout = +15.5 V under load
(+15.6 V in idling mode)
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......