GE Medical Systems
EAGLE 1000
Page 25
Information Technologies
Servicing Instructions
227 468 01 Rev. G
Cuff inflation pressure
adult
160 mmHg
neonate 100 mmHg
Adaptive cuff inflation pressure
After a measurement the inflation pressure of the subsequent measurement depends on
the previous measuring result: it is either 1.8 times the previous mean pressure or 15%
above the previous inflation pressure (whichever is less).
Measurement cycles
Selectable intervals for automatic measurement cycles:
2, 5, 10, 15, 30, 60 minutes
Measuring accuracy
P
sys
/P
dia
<= 8 mmHg, 1st standard deviation
error limits for cuff pressure via the pressure transducer ±3 mmHg
Measurement duration
adult: 90 seconds max.
neonate: 70 seconds max.
typical duration for undisturbed measurement: 30 seconds
Technical inspection, calibration
Technical inspection must be carried out once a year, calibrations every two years (refer to
section 3 Maintenance and technical inspection).
2.1.9 Invasive blood pressure
Signal inputs
2 connections which are both isolated from ground (type CF according to IEC 601-1) for
pressure transducers with ohmic strain-gauge bridge, diagonal resistance of 200 Ohms to
10 kOhms with a sensitivity of 0.005 mV/mmHg per V of supply voltage:
supply voltage for transducer:
5 volts DC
At 50 and 60 Hz:
differential input impedance > 100 kOhms, real component 500 kOhms
common-mode input impedance > 100 kOhms
common-mode rejection > 60 dB up to 60 Hz
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......