Servicing Instructions 227 468 01 Rev.G
3.3 Calibration check of the measuring system
This tests are required in Germany and have to be performed according domestic laws.
The measurement systems for non invasive blood pressure and temperature have to be
checked every 2 years for their accuracy according the following procedure:
3.3.1 Technical Inspections of the Measuring System
The measuring system for noninvasive blood pressure and temperature must be inspected
every two years.
For these inspections, the following devices should be used:
For NIBP digital manometer Diptron 3 plus (measuring range: 500 mmHg)
For TEMP temperature simulator 220 104 01
Conditions:
Please note that the measurement standards used for these inspections
must provide an adequate degree of accuracy. These measurement
standards must be traceable to a national or European standard.
Proceed as follows to perform the technical inspections of the measuring system:
-
NIBP
The conditions stated above must be fulfilled and the functional test for NIBP as per 3.2.5
must be completed.
Step 1: Leakage Text
Step 2: Testing display errors
-
TEMP
The conditions stated above must be fulfilled and the functional test for TEMP as per 3.2.8
must be completed.
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......