Sockets for signal inputs
Circular connectors:
To remove the sockets undo the 2 posidrive plastite screws in each case from the in
side.
When reinstalling each socket insert the wide lug upwards. In the case of the temperature
input socket direct the connections inwards and upwards. First of all, turn the self-
cutting screws in the direction they were turned when being loosened until they click in to
avoid cutting a second thread when being tightened.
For instruments with sockets not attached by screws, the front panel should be replaced
when they are renewed, in this case field replacement is not recommended.
4.5.4 Adjustment procedures
PCB ECG Monitor
ECG ADJUST
OFFSET_1, OFFSET_2:
1 Connect a 5-lead patient cable to the Eagle 1000 and short circuit the ECG inputs.
2 Switch the Eagle 1000 on.
3 Select DISPLAY in the menu (accessed with the Next Menu key). Select Channel 2
ECG and Channel 3 OFF.
4 Select ECG in the menu (accessed with the Prev Menu. Select leads Channel_1 II
and Channel_2 II.
5 Press mm/mV several times so that 5, 10, 20, 5... appear consecutively. Adjust Chan-
nel 1 with
R518
and Channel 2 with
R549
so that when pressing the mm/mV key
several times steps disappear in the zero line on the screen.
GAIN_1, GAIN_2
Have the instrument BIO_TEK Multiparameter Simulator LIONHEART at hand.
1
Connect a 5-lead cable to the Eagle 1000 and the ECG inputs to the LIONHEART
simulator. Switch on the simulator and select ECG AMPLITUDE 1.0 mV. Select
square wave.
2
Switch on the Eagle 1000. Select the configuration menu by pressing the F1 and F5
keys simultaneously. Select muscle filter Off and line filter Off. After selecting Exit
select Apply and save.
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......