1.3.2.2 Data transfer between Eagle and Gateway
Eagle and Gateway will communicate while transferring trend data. Transfer is initiated by
the user request function.
The data are first stored in a file and, when data transfer is complete, arranged in proper
order. The transport monitor waits for the results of this processing procedure.
If an error occurs in the database during this procedure, all the data in the Gateway are
deleted and the entire data must be transferred once more.
When the data transfer is initiated, the time will be sent by the transport monitor real-time
clock. From this the Gateway can identify any possible time misalignment of the internal
clock of the transport monitor and account for this when the trend data are entered.
Sporadic trend data, e.g., non-invasive blood pressure values, are only transferred for the
period of time in which continuous trend data, e.g., ECG, are also available. Earlier or later
parameter readings will be ignored.
1.3.2.3 Connection to the LAN
The QuickLink Station is connected to a Multiport or an SMX using twisted-pair wiring.
The Multiport can lead to the LAN using BNC or AUI wiring.
The SMX 008 leads to the LAN using AUI wiring, the SMX 009 to the LAN using BNC
wiring.
Содержание Eagle 1000
Страница 1: ...Eagle 1000 Patient Monitor Servicing Instructions 227 468 01 SA e Revision G ...
Страница 110: ...GE Medical Systems EAGLE 1000 Page 110 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 112: ...GE Medical Systems EAGLE 1000 Page 112 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 116: ...GE Medical Systems EAGLE 1000 Page 116 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 117: ...GE Medical Systems EAGLE 1000 Page 117 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 118: ...GE Medical Systems EAGLE 1000 Page 118 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 120: ...GE Medical Systems EAGLE 1000 Page 120 Information Technologies Servicing Instructions 227 468 01 Rev G ...
Страница 121: ......
Страница 122: ......
Страница 123: ......
Страница 124: ......
Страница 125: ......
Страница 130: ......
Страница 135: ......
Страница 136: ......
Страница 138: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......