1-18
Dash 2000 Patient Monitor
Revision C
2005873-001
Battery Service
Information Window
Battery Status
Messages
The Battery Service Information window can be accessed using the Trim
Knob in the Monitor Setup Service Mode menu by selecting the Battery
Service softkey. The Battery Service Information window includes the
following information:
BATTERY SERVICE INFORMATION
BATTERY CAPACITY
NEW
ACTUAL FULL
ACTUAL REMAINING
FULL/NEW
REPLACE BELOW
BATTERY TEMPERATURE:
LAST CONDITIONING:
2000
1984
500
99
40
25.0
21-NOV-1998 16:27
mAh
mAh
mAh
%
%
˚
C
009
Battery Service Information Window
The BATTERY LOW message is displayed in the STATUS MESSAGE line
and indicates 10 minutes of battery life remaining. You should connect the
monitor to an AC power source when the message is displayed.
The REPLACE BATTERY message is displayed in the STATUS MESSAGE
line if, when fully charged, the battery reaches only 40% capacity of its
design capacity for cell type.
The BATTERY NEEDS COND system message is displayed when the
battery needs to be reconditioned.
The COND IS RUNNING message is displayed while the battery is being
reconditioned.
The BATTERY DEFECTIVE message is displayed when errors have
occurred within the battery management system or the battery. The reason
of the error can be found in the error logbook (MONITOR SETUP ->
SERVICE MODE -> REVIEW ERRORS -> VIEW OUTPUT ERRORS).
Notify Service.
Environmental Issues: GE Medical Systems
Information Technologies
strives to produce products that are both user safe and environmentally
friendly. We sincerely believe that our products and the production
methods used to produce them, meet these goals.
Battery Notice: This product contains a rechargeable battery. The average
life span of this type of battery is approximately three years. When
replacement becomes necessary, contact a qualified service representative
to perform the replacement.
Disposal Notice: Should this product become damaged beyond repair, or for
some reason its useful life is considered to be at an end, please observe all
local, state, and federal regulations that relate to the disposal of products
that contain lead, batteries, plastics, etc.
Содержание marquette Dash 2000
Страница 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Страница 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 202: ......
Страница 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...