Dash 2000 Patient Monitor
9-13
2005873-001
Smart BP
Pulse Rate
Smart BP is a feature found in ART and FEM Settings menus.
NOTE:
The Smart BP feature is not available when the monitor is set up
for Neonatal-ICU mode.
Smart BP is an arterial (femoral) artifact rejection program which substantially
reduces the occurrences of needless alarms by eliminating most of the alarms
associated with zeroing the transducer, fast flushing the system, and drawing
blood.
When Smart BP is on, and the system recognizes one of these events, the
arterial (femoral) alarms are deactivated and the systolic and diastolic numerics
are replaced with Xs. The message ”ARTIFACT” is displayed. The mean
pressure value is displayed throughout the artifact occurrence.
Safety Features: If zeroing, fast flushing, or drawing blood is not
accomplished within certain time frames, alarms will sound. When artifact
is detected, Smart BP begins to search for the return of a pulsatile
pressure. When 15-20 beats have been detected, numerics are displayed
and alarms are reactivated. If pressure remains below 10 mmHg for more
than 14 seconds, the alarms will reactivate. During sustained high
pressure (drawing blood), you have a maximum of two minutes before
alarms reactivate.
To turn Smart BP on/off, select the SMART BP option from the Settings menu.
This feature can be set in Monitor Defaults.
Another special feature is PULSE RATE. This feature is found in ART, FEM,
and UAC (neonatal mode only) pressure sites. When turned on, a Rate value
is displayed in the appropriate parameter values window.
ART
200/ 80
S mmHg
RT.
70
72
94
136/
Pulse
Rate on
106
Pulse Rate Displayed in ART Parameter Window
To turn the pulse rate value on and off, simply select PULSE RATE in the
applicable Settings menu.
Содержание marquette Dash 2000
Страница 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Страница 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 202: ......
Страница 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...