Revision C
Dash 2000 Patient Monitor
9-15
2005873-001
Troubleshooting
Problem:
Displayed pressure values
are different than expected
Problem:
Smart BP is on. Artifact is
sensed without flush, draw,
or zero
Solution
:
Check tubing for bubbles.
Remove excessive tubing.
Check phlebostatic axis placement of transducer.
Rezero pressure:
1. Place the transducer at the left atrial level.
2. Close transducer stopcock to patient.
3. Open the venting stopcock to air.
4. Press FUNCTION key on the front panel of the monitor.
5. Close the venting stopcock to air.
6. Open transducer stopcock to patient.
Solution:
Turn Smart BP OFF then ON. If problem persists, you may need to disable
Smart BP by turning it off for that pressure.
Use Alarm Pause feature prior to drawing blood to reduce unnecessary alarms
if Smart BP is disabled.
Содержание marquette Dash 2000
Страница 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Страница 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 202: ......
Страница 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...