9-10
Dash 2000 Patient Monitor
Revision C
2005873-001
Change Name
Zero
This option allows you to change the designated name for the pressure
connector. Remember, the monitor has preset the name for the pressure
connector. (Refer to this chapter, Assigned BP Names.)
Having the name properly reflect the site is important for proper processing of
the waveform since different algorithms are used for different pressure sites.
Follow this procedure to change a name:
1. Select CHANGE NAME option from the pressure menu. A popup menu
opens showing all choices. The name presently assigned is designated
with the pointer (>).
MAIN
MENU
CHANGE
NAME
ART
ART
FEM
PA
CVP
RA
LA
ICP
SP
>
103
CHANGE NAME Popup Menu
2. Rotate the Trim Knob control to place the pointer in front of the desired
name.
The change will not take effect until the popup menu is closed.
3. Press the Trim Knob control to confirm the change and close the popup
menu. This causes a change in the parameter window and the Main
Menu is displayed.
Use the ZERO option to zero the transducer.
1. Place the transducer at the left atrial level.
2. Close the transducer stopcock to the patient.
3. Open the venting stopcock to air (atmosphere).
4. Use the Trim Knob control to select (highlight and press) the ZERO
option in the pressure menu.
5. Verify that zero reference has been established. (Watch the pressure
parameter window messages.)
6. Close the venting stopcock to air (atmosphere).
7. Open the transducer stopcock to the patient. Within seconds pressure
numerics should be displayed in the pressure parameter window.
Содержание marquette Dash 2000
Страница 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Страница 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 202: ......
Страница 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...