13-6
Dash 2000 Patient Monitor
Revision C
2005873-001
Troubleshooting
Messages
If you experience some problems with temperature monitoring, one of the
following messages may be displayed in the TP parameter window:
CHECK
There is a 0.1 oC deviation between the temperature value and the
internal calibration. No temperature value will be displayed. Service
on the monitor is required.
CAL FAIL
There is a 1 oC deviation between the temperature value sensed and
the internal calibration. No temperature value will be displayed.
Service on the monitor is required.
SENSOR
The temperature sensor has failed.