Revision C
Dash 2000 Patient Monitor
8-5
2005873-001
3-Leadwire Electrode
Placement
When a 5-leadwire electrode configuration is not desirable, a 3-leadwire
electrode configuration can be used.
348A
RA
LA
LL
348B
R
L
F
AHA Configuration
IEC Configuration
Three-leadwire
Configuration
NOTE: Electrode configuration will vary depending on the type of
leadwire set you are using. Refer to ”Three-leadwire
Configuration” below.
Right arm and left arm electrodes should be placed just below the right and left
clavicle.
Left leg electrode should be placed on a non-muscular surface on the lower
edge of the rib cage.
The molded 3-leadwire sets can be placed in the 5-lead Multi-Link patient
cable.
Standard Molded 3-leadwire Set (Rotating Reference)
•
Selectable lead I, II, or III cable with a rotating reference (right arm,
left arm, left leg).
Operation of the monitor with this standard cable allows you to select one
of three leads (I, II, or III) for monitoring.
When using the standard, 3-leadwire configuration, the following operating
conditions occur:
•
LD ANALYSIS automatically switches to single lead analysis. If an
attempt is made to change to multi-lead analysis, a message "NOT
POSSIBLE WITH 3 LEAD CABLE" will appear briefly and no change
will occur.
•
DISPLAY LEAD choices are limited to I, II, and III.
•
Any options usually allowing more than one ECG lead selection will be
disallowed. A message line is displayed briefly, indicating such.
Содержание marquette Dash 2000
Страница 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Страница 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Страница 202: ......
Страница 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...