Revision C
Dash 2000 Patient Monitor
8-7
2005873-001
Electrode Placement
for Pacemaker
Patients
Electrodes need to be repositioned to modify detection of the electrical signals
generated by the pacemaker. Following is a suggested configuration:
NOTE:
When using this configuration, display Lead II as your primary
ECG lead.
AHA Configuration
IEC Configuration
The right arm electrode is moved down to the 5th intercostal space, and the left
leg electrode is moved up to the 5th intercostal space.
NOTE:
After all electrodes are in place, ensure minimum of 1/2 mV of
signal is present on each lead (I, II, III, V).