4-13
Chapter 4 Connection to the External Devices
PC Communication
PC Communication
This section explains about the PC communication setup procedure.
By using the PC communication function, vital data measured on the bedside monitor can be transmitted to PC.
Connection with the System
1
Connect the accessory cable to the serial connector (COM) on the AC Unit (DSA-81).
External Device Setup
To transmit the data to PC, "External Device" setup is required.
1
Press the [Menu], [Setup], [Initial Settings], [External
Device], [AC Unit] keys.
The screen to set the connecting device type for
each port will be displayed.
2
Select the serial connector (COM) which the PC was
connected.
3
Press the [Other] key.
4
Press the [PC Comm.] key.
NOTE
If the PC communication is disconnected, <Check System Conn.> will be displayed.
3
4
2
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......