3-3
Chapter 3 Using the External Media
Data Backup/Copy Using the USB Memory/SD Card
2
Press and hold the [Start] key.
3
When the process completes, remove the SD card.
4
Press the [Yes] key if OK to read the data from the USB memory/SD card.
With this process, the patient data, setup data, full disclosure waveform data can be transferred.
CAUTION
While copying the data, do not disconnect the DSA-81. It may damage the setup data.
To copy the setup data saved on the AC Unit (DSA-81), connect the AC Unit (DSA-81)
to the DS-8007 before copying the data.
If not copying the setup data saved on the AC Unit (DSA-81), disconnect the AC Unit
(DSA-81) from the DS-8007 before copying the data.
Make sure to power cycle the system after the data is read from the SD card. By power
cycling the system, the read data will become effective.
Displayed Messages on the External Media Menu
For "External Media", the following messages will be displayed depending on the external media status.
Data that can be Backed Up/Copied
The setup data such as monitoring condition, alarm setting, and patient data such as graphic trend and tabular trend
can be backed up/copied.
*Backup: By selecting [Write], setup data and patient data will be both backed up.
*Copy: By selecting [All Data], only the setup data saved on the SD card or USB memory will be copied to the
patient monitor.
NOTE
Selecting [All Data] will not copy the patient data.To transfer the patient data, use the
"Transfer Full Disclosure Waveform Card" function.
While the data is being backed up or copied, full disclosure waveform data will not be saved.
On the USB memory, full disclosure waveform data cannot be saved.
Message
Description
Full Disclosure Waveform Card
The card is formatted on the used DS-8007.
Full Disc. Wave Card (Other Unit)
The card is formatted on other DS-8007.
Format the card.
The card is not formatted on the DS-8007.
There is no media.
The media is not inserted.
For data transfer
USB memory or unspecified SD card
Setup Data
Data
Description
Parameter Setup
Stores the monitoring condition (size, lead, etc.) for all the monitoring
parameters.
Alarm
Stores the alarm threshold level.
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......