Chapter 9 Maintenance Check
Handling and Storage of Lithium-Ion Battery Pack (BTO-008)
Handling and Storage of Lithium-Ion Battery Pack (BTO-008)
This section describes the handling and storage of the BTO-008 battery pack. Refer also to the BTO-008 Operation
Manual.
Handling the Battery
For uninterrupted monitoring, charge the battery when the battery level is low.
When the battery operation time becomes short even after it is fully charged, the battery needs to be replaced.
The battery should be charged at room temperature (10°C to 30°C).
The lithium-ion battery can only be charged in the specified operating temperatures of the equipment. Refer to
the operation manual of the lithium-ion battery (BTO-008) for details.
When using the battery for the first time, or using after leaving it for a while, make sure to charge the battery
before use.
Storing
To take advantage of the characteristic of the battery pack, pay attention to the following when storing.
Storage Temperature and Humidity for the Battery
Store in an environment specified below without corrosive gas.
Do not store in an environment outside the specified temperature range or excessive high humidity. This may
result in leakage caused by expansion/contraction inside the battery or rusting of the metal part.
Long-Term Storage
If the battery is left installed in the monitor without use for a long period of time, the capacity recovery after
storage may be degraded.
When storing the monitor for a long period, remove the battery from the monitor.
05
Patient Leakage
Current (SFC)
Caused by an external
voltage on the patient
connection of an F-
TYPE APPLIED
PART.
Measure the patient leakage current under single
failure condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
Patient Leakage Current (SFC) 0.05 mA
or less
(CF Type)
06
Patient Auxiliary
Current (NC)
Measure the patient auxiliary current (NC) under
normal condition using a leak measurement safety
tester.
Test according to the test method of IEC 60601-1.
Patient Auxiliary Current (AC) 0.01 mA or
less
(CF Type)
07
Patient Auxiliary
Current (SFC)
Measure the patient auxiliary current (SFC) under
single failure condition using a leak measurement
safety tester.
Test according to the test method of IEC 60601-1.
Patient Auxiliary Current (SFC) 0.05 mA or
less
(CF Type)
Storage Period
Storage Temperature
Storage Humidity
Within 30 days
-20°C to 60°C
20% to 85% (non-condensing)
Within 90 days
-20°C to 45°C
Within 1 year
-20°C to 20°C
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......