5-16
Chapter 5 Initial Settings
User I/F
5
The backup status when the power is turned ON and when the patient is discharged can be set for each item.
[Backup]: The setting will be backed up.
[Initialize]: Initializes the settings. The initialized settings are as follows.
CAUTION
When the discharge process is performed, patient data such as recall and trend will be
initialized. The parameter and alarm settings will be reset according to the settings made
under [Menu>Setup>Initial Settings>User I/F>Power ON/Discharge).
When the discharge process is performed on the central monitor, alarm will be reset
according to the settings made under "Admit Setup" of the central monitor.
NOTE
The operation after the power is turned ON will be according to the setting made on [Initial
Settings] > [User I/F] > [Power ON/Discharge]. However, if the power was turned OFF for
less than 30 seconds, the setting before the power was turned OFF will remain.
If the "Monitor Mode" setting is other than [Backup], only the display configuration and
pacemaker settings can be performed.
6
When "Link with Patient Class." is set to [ON], the operation will differ depending on the "Monitor Mode" setting
under "Power ON/Discharge". (
"To Program the User Mode" P5-23)
If [Backup] is set for "Monitor Mode" under "Power ON/Discharge", the settings will be as follows.
Selection other than Backup
Item
Setup
Power ON/Discharge
Monitor Mode
Current Mode
Monitor Mode 1 to 6
The setting will be initialized to the selected mode.
Display Configuration
Display Mode 1 to 6
The setting will be initialized to the selected mode.
Patient Classification
Adult, Child, Neonate
The setting will be initialized to the selected patient
classification.
Pacemaker
Not Used
"Not Used" will be set for "Pacemaker".
Alarm Settings
Initialize
The setting will be initialized with the currently selected
mode.
ECG1, ECG2 Lead
Initialize
The setting will be initialized with the currently selected
mode.
ECG1, ECG2 Size
Initialize
The setting will be initialized with the currently selected
mode.
Impedance
Mode ON/OFF
Initialize
The setting will be initialized with the currently selected
mode.
CVA Detect
OFF
CVA detection will be set to OFF.
NIBP Auto Mode
OFF
NIBP auto mode will be turned OFF.
OFF, 2.5 min
If NIBP Auto Mode is OFF, 2.5 min. interval will be set.
OFF, 5 min
If NIBP Auto Mode is OFF, 5 min. interval will be set.
2.5 min.
NIBP auto mode will be set to 2.5 min. interval.
5 min.
NIBP auto mode will be set to 5 min. interval.
BP Scale
Initialize
The setting will be initialized with the currently selected
mode.
SpO
2
Averaging
Initialize
The setting will be initialized with the currently selected
mode.
CO
2
Scale
Initialize
The setting will be initialized with the currently selected
mode.
EtCO
2
Peak Duration
10 sec.
EtCO
2
peak picking duration will be set to 10 sec.
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......