2-5
Chapter 2 Network System Construction
Wired Network System
When the BP measurement unit is "kPa", BP waveform, BP numeric data, NIBP numeric
data, NIBP list will not be transmitted. It will be treated as not measured data, and will not
be displayed on the central monitor. Also, the alarm limit setup from the central monitor
cannot be performed.
The following arrhythmia alarms cannot be transmitted.
TACHY, BRADY, COUPLET, PAUSE, TRIGEMINY, TRIPLET, EXT TACHY, EXT
BRADY, R on T, MULTIFORM, VENT RHYTHM, SVT, IRREGULAR RR, PROLONGED
RR, S FREQUENT, S COUPLET, VPC, SVPC, PACER NOT CAPTURE, PACE NOT
PACING
Arrhythmia alarm of "Slow_VT" will be transmitted as "VT" .
On a wired network, waveform, numeric data, and alarm of TEMP3 to 6 will not be
transmitted. Also, the displayable waveform, numeric data, alarm differs depending on
the connected central monitor. Refer also to the operation manual for the respective
central monitor.
The numeric data and alarm of PR_IBP will not be transmitted to the central monitor.
Even if the PR_IBP alarm is generated on the DS-8007, this alarm will not be generated
on the central monitor.
When the DS-5800N/NX/NX
MB
is used as a central monitor, recall, graphic trend, and
tabular trend will not be displayed, and
Σ
recording cannot be performed. For the ST
display, overlap waveform will not be displayed on the DS-5800N/NX/NX
MB
until 15
minutes have passed since the reference waveform is set on this monitor.
The ST display on the central monitor will be distorted when the ECG lead (ECG1 or ECG
2) is changed on this monitor. Redrawing the ST display will return the display to normal.
When connected to the DS-LAN II network, data uploading for the MPDR function is not
possible.
NOTE
If the numeric data is displayed as "xxx" (out of measurement range) on this monitor,
maximum or minimum value of measurable range will be transmitted to the central monitor.
Value Outside the
Measurement Range
Central Monitor Display
HR
301 bpm and above
300 bpm
Respiration Rate
151 Bpm and above
150 Bpm
Blood Pressure
–51 mmHg and below
301 mmHg and above
-50 mmHg
300 mmHg
Temperature
-0.1°C and below
45.1°C and above
0°C
45.0°C
Pulse Rate
(Masimo Unit)
240 bpm and above
25 bpm and below
239 bpm
26bpm
Pulse Rate
(Nellcor Unit)
251 bpm and above
300 bpm
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......