1-8
Chapter 1 Installation of the Unit
Power Source and Ground Connection
6 AUX Connection Cable (CJO-25TR)
Power Source and Ground Connection
This section explains about the power connection.
Power Connection of the Main Unit
1
Connect the power cable to the main unit (DS-8007).
1
Connect the power cable (CS-18) to the rear side of the AC Unit (DSA-81).
2
Use the lever to lock the cable retainer clip.
3
Connect the power cable to a hospital grade outlet (3-pin grounded outlet).
To disconnect the power cable, unplug one end from the outlet, and the other end from the rear side of the main unit
after releasing the lock lever.
2
Turn ON the standby switch on the main unit.
AC power will be supplied and the power supply LED on the
front side of the main unit will light.
1
Power Supply LED
Green: In normal operation
Orange: Standby Mode
Light Off: During battery operation
(The AC power cable is not connected.)
2
Battery Charging LED
Green: Charging is complete
Orange: Charging is in process
Light Off: During battery operation, or when battery is
not installed, or when battery charging is ceased (due
to temperature, etc.)
Flash: Battery charging error
NOTE
If the battery charging LED flashes, battery charging error is occurring. Remove the battery
and install it again. If the error persists, contact your nearest service representative.
1
2
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......