5-17
Chapter 5 Initial Settings
User I/F
When "Link with Patient Class." is set to [ON], it is recommended to set [Initialize] for "Alarm" under "Power
ON/Discharge".
If settings other than [Backup] is set for "Monitor Mode" under "Power ON/Discharge", the selected monitor
mode settings will be applied regardless of the ON/OFF setting of "Link with Patient Class.".
Shortcut Menu
The keys to be displayed on the menu display can be customized.
By registering the frequently used menu, quick access to the menu will become possible.
5 shortcut menus can be set.
Power ON/Discharge
Operation
Patient
Classification
Alarm Settings
Backup
Backup
Patient
Classification:
Current patient classification will be maintained.
Alarm Settings:
Current alarm settings will be maintained.
Initialize
Patient
Classification:
Current patient classification will be maintained.
Alarm Settings:
Alarm settings of the monitor mode selected for "Link with Patient
Class."
Adult
Backup
Patient
Classification:
Patient classification will be set to "Adult".
Alarm Settings:
Current alarm settings will be maintained.
Initialize
Patient
Classification:
Patient classification will be set to "Adult".
Alarm Settings:
Alarm settings of the monitor mode selected for "Link with Patient
Class."
Child
Backup
Patient
Classification:
Patient classification will be set to "Child".
Alarm Settings:
Current alarm settings will be maintained.
Initialize
Patient
Classification:
Patient classification will be set to "Child".
Alarm Settings:
Alarm settings of the monitor mode selected for "Link with Patient
Class."
Neonate
Backup
Patient
Classification:
Patient classification will be set to "Neonate".
Alarm Settings:
Current alarm settings will be maintained.
Initialize
Patient
Classification:
Patient classification will be set to "Neonate".
Alarm Settings:
Alarm settings of the monitor mode selected for "Link with Patient
Class."
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......