5-6
Chapter 5 Initial Settings
Alarm Related Setup
[OFF]: Arrhythmia analysis will not be suspended even when a noise is continuously interfering.
CAUTION
When "Suspend Arrhy. Analysis during Noise Interference" is set to [ON], and the
suspended duration continues for more than 30 seconds, "Cannot analyze" message will
generate.
6
Set the "Auto Alarm Setup".
[Enable]: [Auto] key will be displayed on the alarm setup menu.
[Disable]: [Auto] key will not be displayed on the alarm setup menu.
7
Set the "Lower Limit for Alarm Volume".
Set the lower limit of alarm volume for "Vital Alarm", "Ventilator Alarm", "Status Alarm".
WARNING
Changing the setting for "Alarm System" will also change the alarm volume and tone
setting. As the "Lower Limit for Alarm Volume" may also change, make sure to check the
volume and tone on the "Tone/Volume" setup menu.
The lower limit of adjustable alarm volume range on the "Tone/Volume" setup menu will be set. The lower
limit level can be set according to the alarm level priority, Urgent>Caution>Status.
[Test]: The test sound will be generated with the set volume.
8
Alarm Mute
[ON]: The alarm mute function will turn ON.
[OFF]: The alarm mute function will turn OFF.
9
Alarm Mute Reminder
A reminder message/sound can be displayed/generated after preprogrammed duration to remind the alarm
mute condition.
Select from [15 min.] / [30 min.] / [60 min.] / [120 min.].
[OFF]: A reminder message/sound will not be displayed/generated.
10
Set the "HR/PR Lower Limit during Alarm Auto Setting".
[OFF]: No limit will be set.
[30 bpm]: When the auto alarm is set and the lower limit is below 30 bpm, the lower limit will be fixed to 30
bpm.
[40 bpm]: When the auto alarm is set and the lower limit is below 40 bpm, the lower limit will be fixed to 40
bpm.
11
Set the "Alarm Threshold Limit".
The alarm threshold range for each parameter can be set.
For the parameter set to [Enable], the alarm threshold level outside the set range cannot be set.
12
Set the "Alarm Level".The alarm level for numeric data alarm, arrhythmia alarm, technical alarm can be set.
The alarm level can be selected from S, H, M, L according to the priority. ("S" is the highest priority alarm.)
1
Select the alarm level from [DS-LAN Standard Setup] or [User Setup].
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......