9-2
Chapter 9 Maintenance Check
Daily and Periodic Check
Daily Check Items and Procedures
Perform the daily check using the "Daily Check List".
The "Daily Check List" is on the last page of this manual.
If the equipment fails any check item on the daily check list, the general judgment will be "Fail". Repair the
equipment so that it passes all the check items.
Use the equipment only if the judgments for all the items are "OK".
No.
Check Items
Check Procedure
Criteria
1. Cleanness, Cleaning (Note: Turn OFF the power, and disconnect the power cable and battery before cleaning.)
01
Cleanness
If contamination is found, perform proper
disinfection before the daily check.
No contamination should be found.
02
Cleaning
Visually check the exterior, and perform
proper cleaning.
For details, refer to "Chapter 8 Cleaning/
Disinfecting/Storing" of Maintenance Manual.
It should be clean.
No.
Check Items
Check Procedure
Criteria
2. External Appearance
01
External
Appearance
Visually check the exterior for scratches,
cracks, and rust.
No abnormality should be found.
02
Cables
Check that the cables are intact and firmly
connected.
The cables should be intact and firmly connected.
03
Installation
Check whether the equipment is installed on a
level surface.
The installation area must be level and free from
vibration and shock.
04
Installation
Check whether the equipment is installed in a
place susceptible to adverse environment.
The temperature and humidity of the installation
area must be as specified.
The equipment should not be subjected to
splashing water or chemicals.
No.
Check Items
Check Procedure
Criteria
3. Operation
01
Function
Turn ON the power, and check that the
equipment operates normally.
The home display should appear, and the power
supply LED should light.
The date and time should be correct.
02
Function
Turn ON the power, and check that the
equipment operates normally.
With the BP relay cable and BP transducer
connected, pressing the BP zero balance switch
should start the zero balance.
Pressing the NIBP Start/Stop key should inflate the
NIBP cuff.
Connecting the SpO
2
sensor should light the
sensor LED.
03
Function
(When HPD-810/820, HCP-810/820 are used)
The home display should appear, and the power
supply LED should light in green.
When the sampling tube is connected, "0" should
be displayed in the numeric data box.
04
CO
2
Calibration
(When HCP-810/
820 is used)
Check the date of the previous calibration.
(Refer to the following caution.)
Should be within one year.
Check the remaining time until the next
calibration.
[Menu] > [Parameter] > [CO
2
]> [CO
2
Cal.]
Should not be 0 hrs.
05
Alarm Indicator
Check the alarm indicator operation by
pressing the [Indicator Test] key.
It should light with the color of each level.
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......