8-4
Chapter 8 Cleaning/Disinfecting/Storing
Cleaning the Equipment and Sensors
BP Transducer
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
Do not reuse / re-sterilize the disposable type transducers.
SpO
2
Sensor
Disinfect the SpO
2
sensor according to the manufacturer's guidelines. Do not reuse/resterilize the disposable SpO
2
sensor.
Nellcor Sensor
Do not soak the sensor in water or antiseptic solution.
Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not sterilize by irradiation, steam, or
ethylene oxide.
OxiMax is a single-patient use type sensor. Do not reuse or resterilize.
Masimo Sensor
Do not soak the sensor or patient cable in water or antiseptic solution.
(Sensors and connectors are not water-proof.)
Do not sterilize the sensors and cables by irradiation, steam, or ethylene oxide.
The Masimo disposable sensor can be reused on the same patient if the light emitting and receiving part is
clean, and if it is still adhesive to the skin.
For most of the sensors, the adhesiveness will return by cleaning the sensor with alcohol and completely drying
it before applying it to the patient. For details of the cleaning procedure, refer to the instruction manual of the
sensor.
Disinfect the Masimo reusable sensor and patient cable according to the manufacturer's guidelines.
When cleaning the Masimo reusable sensor and patient cable, disconnect them from the main unit, and follow
the procedure below.
1
Wipe the sensor and cable using 70% isopropyl alcohol cotton.
2
Dry it completely with air before reusing.
Temperature Probe
Disinfect the temperature probe according to the manufacturer's guidelines.
When cleaning the relay cable, follow the procedure below.
1
Wipe the cable using 70% isopropyl alcohol cotton.
2
Dry it completely with air before reusing.
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......