2-4
Chapter 2 Network System Construction
Wired Network System
this monitor and the central monitor. The hemodynamic data edited on the central monitor will be deleted.
The hemodynamic data edited on this monitor will be transmitted to the central monitor.
6
Set the "CO
2
(mmHg) Upper Limit of Transmission".
If the CO
2
measurement unit is "mmHg", and the CO
2
value is 100 mmHg or above, whether or not to limit the
value for transmission to the central monitor can be set.
[No limit]: Actual CO
2
value will be transmitted to the central monitor even if the value is 100 mmHg or
above.
[99mmHg]: 99 mmHg will be transmitted as the CO
2
value if the value is 100 mmHg or above.
7
Press the [Menu], [Initial Settings], [System], [Other] keys, and set the "Search Patient ID". (
[Enable]: Patient data can be searched on the patient data server using the patient ID.
(
Operation Manual "Entering Patient Information from the Patient Data Server (When DS-LANIII is
used)" P5-4)
[Disable]: Patient data will not be searched on the patient data server.
Precautions about Printing/Display
CAUTION
Precautions Common for the DS-LAN II/DS-LAN III Network
If the "RR/APNEA Alarm Source" setting is other than [Impedance] (or, if [Auto] selects a
setting other than [Impedance]), the RESP waveform will not be transmitted on a wired
network.
If the "RR Alarm/APNEA Source" setting is other than [CO
2
] (Or, if [Auto] selects a setting
other than [CO
2
]), the CO
2
waveform will not be transmitted on a wired network.
For the numeric data displayed as "xxx", maximum or minimum value of measurable
range will be transmitted.
The numeric data displayed as "--- " will be treated as not measured data.
If the measurement unit of CO
2
concentration is mmHg and [99 mmHg] is selected for
"Upper Limit of CO
2
(mmHg) Transmission" under [Initial Settings>System>DS-LAN or
Telemeter), the CO
2
value of 100 mmHg or above will be transmitted as 99 mmHg.
As this monitor do not have the arrhythmia template display and 12-lead ST display
function, these display on the central monitor will not be corresponded.
When connected to a wired network, the time/date will synchronize with the central
monitor. In this case, the time/date cannot be changed on this monitor.
On the central monitor, the respiration waveform and RR value based on the "RR/APNEA
Alarm Source" selected on this monitor will be displayed. The same parameter for the RR
and apnea will be monitored on this monitor and the central monitor.
The alarms that can be notified to the central monitor depend on the model type and
software version of the central monitor. For details, refer to the operation manual of the
central monitor.
CAUTION
Precautions about the DS-LAN II Network
The BP measurement unit should be set to "mmHg".
The data cannot be output to the AU-5500N.
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......